td7200 calibrator for electrosurgical analyzers -必赢亚赢

for more product details, please call us:0731-84930888

home> electrical measurement> medical calibrators

td7200 calibrator for electrosurgical analyzers

专用于校准商用的高频电刀分析仪

参考标准:jjf 1217-2009 《高频电刀校准规范》


采用恒压源模式校准高频电刀分析仪的漏电流及功率,并检测其线性度和频响

内置测量标准的输入带宽优于 30 mhz @ 3db

overview

model td7200, a calibrator for commercial electrosurgical analyzers. the apparatus can simulate electric coagulation, electric cutting, mixing and other functions of high-frequency electric knives. it can accurately output high frequency voltage, current, power and other electrical parameter signals to calibrate the electrosurgical analyzers.

features

● the functions of outputs and measurements are integrated with small size and light weight, and can be used in field and laboratory.

● calibrate the leakage current and power as well as their linearity and frequency response of high frequency electrosurgical analyzers by the constant voltage source mode.

● the input bandwidth of the built-in measurement standard is better than 30 mhz @ 3 db.

● proprietary software (option): support the calibration by semi-automatic or full-automatic, data management and report export.

electrical specifications

111

general specifications

● line power: 220 v ±22 v, 50 hz ±2 hz

● maximum power consumption: 2.5 kva

● operating environment: 0 °c to +40 °c, 20 % to 85 % r·h, non-condensing

● storage environment: -20 °c to +70 °c, <85 % r·h, non-condensing

● communication: rs-232 interface



powered by © 2011-2013 inc.
网站地图