th0400 precision coaxial shunt

适用于测量精密交直流电流,校准交直流电流表、精密电流标准源等设备

共含 14 只不同量程的精密同轴分流器 

电流测量: dc ~ 100 khz,0.1 ma ~ 110 a 

标称电压值均为 1 v

最优准确度达 20 ppm

overview

model th0400, a series of 14 various ranges precision coaxial shunts, applied to measure ac/dc current from 0.1 ma to 110 a and frequency from dc to 100 khz, the optimal accuracy is up to ±20 ppm.

applied to measure precision ac and dc current and calibrate precision ammeters and standard current sources.

features

● the nominal voltage is 1 v, can measure current by directly matching high-precision voltmeters or other measurement instruments.

●  for 1 ma, 10 ma and other high resistance current shunts, combine with the model th0600 to avoid errors caused by wire connection or input impedance drops of digital meters at high frequency.

● compared to 0.5 v, 0.8 v and other voltage value, the 1v nominal voltage has a higher noise ratio, easier current conversion, faster error calculation, and accords with the 1-2-5 sequence principle of modern measurement.

● precision components and production processes ensure the resistance value stable, the typical resistance change value is ±5 μω/ω in 1 year, and the maximum resistance change value is ±16 μω/ω in 1 year.

● good heat dissipation, lower temperature coefficient and self-heating power factor ensure a good stability.

● small phase displacement between the input current and the sampled voltage, the phase displacement is less than 0.0015 ° even at 1 khz. this feature is critical for the measurement of non-sinusoidal and wideband signals, applied to analyze power quality and calibrate digital power meters.

● the capacity of loading current and withstanding current impulse is limited, but you can use a set of shunts to obtain a wide range of current measurement.

electrical specifications

resistance

1

current

2

phase

3

general specifications

● operating environment: +13 °c to 33 °c, r·h ≤50 %, altitude from 0 to 3,000 m.

● calibrate temperature: +18 °c to 28 °c

● storage environment: +5 °c to +45 °c, r·h from 15 % to 80 %.

● input interfaces: type-n (f) for 1 ma to 20 a, type-lc (f) for 50 a to 100 a.

● output interfaces: type-n (f)

● notes: the stability of resistance is affected by humidity, but the change is reversible. when stored in a harsh environment, it may cause the resistance to change by a value of ±20 ppm,  and return to the original resistance within the next 30 days if the coaxial shunts are stored or operated in the specified environment later.